高圧電子顕微鏡法


Features of the ultra-high voltage electron microscope (H-3000)

photo

Figure 1 Ultra-High Voltage Electron Microscope - Hitachi H-3000 in Osaka University


The 3 MV ultra-high voltage electron microscope (Hitachi H-3000) was developed based on various advanced technologies, such as the high stability of the high voltage generating circuit, the reduction of power losses, the improvement of the response speed of the objective lens, negative ion removal equipment, a remote control system, and an online image enhancement system. The ultra-high voltage electron microscope is widely used by researchers in many fields, including materials science, nanotechnology, biology, and medical science. The particular advantages of the H-3000 can be summarized as follows:

  1. Remarkable increases in the maximum observable thickness of a specimen.
  2. Various types of in-situ observations made possible by the very large specimen space.
  3. Lattice defect introduction and/or non-equilibrium phase formation through the interaction between high energy electrons and constituent elements in materials.
    Main specifications
  • High voltage generator
    • Symmetrical Cockcroft–Walton circuit: 35-stage
    • Electricity consumption in high voltage tank: under 1 kW
    • Acceleration voltage:
      • Normal: 3.0 MV (3000 kV)
      • Medium: 0.5, 1.0, 1.5, 2.0, 2.5 MV
      • Maximum: 3.5 MV
      • Stability of high voltage: less than 2.0×10-6/min
  • Electron gun and acceleration tube
    • Cathode: LaB6 single crystal (thermal type)
    • Beam current: maximum 20 mA
    • Acceleration tube: 22 kV/step × 138 steps (3 MV)
    • Insulating gas: SF6 (4 atom)
  • Irradiation lens system
    • Ion trapping system: electron beam shift by polarization coil
    • Lens constitution: double-condenser lens
    • Beam angle: less than 10-3
  • Imaging lens system
    • Lens configuration: six-step lens system (no image rotation)
    • Magnification: 200 to 1,000,000 (30-step switching)
    • Camera length at electron diffraction: maximum 14 m (six-step switching)
    • Resolution: less than 10 nm
  • Objective lens
    • Magnetomotive force: maximum 23 A/V1/2 (3 MV)
    • Focusing length: less than 11 mm
    • Spherical aberration coefficient: less than 10 mm
    • Chromatic aberration coefficient: less than 10 nm
    • Defocused point: minimum 5 nm
  • Specimen space
    • Holder: Top-entry type and side-entry type
    • Orientation angle of top-entry type: ±30 degrees (double-tile type)
    • Orientation angle of side-entry type: ±45 degrees (double-tile type)
    • Specimen motion: ±1 mm (mechanical system), ±1.5 mm (electrical system)
    • Vacuum: 5×10-6 Pa (oil-free vacuum system)
  • Camera room and operation room
    • Photo system: imaging plate (32 sheets × 2), film (50 sheets × 2)
    • Fluorescent substance: P22, YAG
    • Fluorescent screen: maximum 60 × 80 mm
  • TV camera
    • High resolution type: 1125-line HARP camera
    • High sensitivity type: slow-scan CCD camera with cooling
    • Video recording: NTSC mode HARP camera
  • Remote control system
    • Operation system: Console panel for remote operation
    • Controlling system: main control workstation system and decentralized data processing system
    • Monitoring system: X-ray, oxygen deficiency, beam position, specimen holder


Cross section for atomic displacement by fast electron, (calculated by the McKinley-Feshbach formula)

Table 1
Threshold energy (Ed), threshold acceleration voltage (Ud), and cross section (s) for atomic displacement of pure element by fast electron irradiation [1, 2]. Most of the values for Ed and Ud were taken from the references [3-5]. A list of the systematic experimental data was firstly summarized by K. Urban [3,4] and slightly modified by H. Fujita [5]. The s values were calculated using the McKinley-Feshbach formula, which is frequently used in radiation damage analysis [6-9]., (Updated on 02/21/2014)

photo


012 Mg (Atomic number 12, Atomic mass number 24.31, Ed 14 eV)

photo

013 Al (Atomic number 13, Atomic mass number 26.98, Ed 14 eV)

photo

014 Si (Atomic number 14, Atomic mass number 28.09, Ed 19 or 13 eV)

photo

022 Ti (Atomic number 22, Atomic mass number 47.88, Ed 19 eV)

photo

023 V (Atomic number 23, Atomic mass number 50.94, Ed 26 eV)

photo


024 Cr (Atomic number 24, Atomic mass number 51.9961, Ed 26 eV)

photo


026 Fe (Atomic number 26, Atomic mass number 58.93, Ed 22 eV)

photo


027 Co (Atomic number 27, Atomic mass number 58.93, Ed 22 eV)

photo


028 Ni (Atomic number 28, Atomic mass number 58.69, Ed 24 eV)

photo


029 Cu (Atomic number 29, Atomic mass number 63.55, Ed 19 eV)

photo


040 Zr (Atomic number 40, Atomic mass number 91.22, Ed 26 eV)

photo


041 Nb (Atomic number 41, Atomic mass number 92.21, Ed 33.5 eV)

photo


042 Mo (Atomic number 42, Atomic mass number 95.94, Ed 30 eV)

photo


046 Pd (Atomic number 46, Atomic mass number 106.4, Ed 34 eV)

photo


047 Ag (Atomic number 47, Atomic mass number 107.9, Ed 24 eV)

photo


050 Sn (Atomic number 50, Atomic mass number 118.7, Ed 12 eV)

photo


051 Sb (Atomic number 51, Atomic mass number 121.75, Ed 14.6 eV)

photo


060 Nd (Atomic number 60, Atomic mass number 144.2, Ed 19 eV)

photo


072 Hf (Atomic number 72, Atomic mass number 178.5, Ed 29 eV)

photo


073 Ta (Atomic number 73, Atomic mass number 180.9, Ed 32 eV)

photo


074 W (Atomic number 74, Atomic mass number 183.84, Ed 35 eV)

photo


078 Pt (Atomic number 78, Atomic mass number 195.1, Ed 33 eV)

photo


079 Au (Atomic number 79, Atomic mass number 197, Ed 33 eV)

photo

Figure 1
Cross section (s) for atomic displacement of pure element by fast electron irradiation. Most of the values for Ed and Ud were taken from the references [3-5]. The s values were calculated using the McKinley-Feshbach formula, which is frequently used in radiation damage analysis [6-9]., (Updated on 02/20/2014)

REFERENCE
[1] T. Nagase, Advanced materials design by irradiation of high energy particles, in: Progress in Advanced Structural and Functional Materials Design, Ed., T. Kakeshita, Springer, 2013. pp. 137-153., ISBN 978-4-431-54063-2, http://www.springer.com/materials/structural+materials/book/978-4-431-54063-2
[2] T. Nagase, T. Sanda, A. Nino, W. Qin, H. Yasuda, H. Mori, Y. Umakoshi, J.A. Szpunar, J. of Non-Cryst. Solids, 358, 502-518 (2012)., http://dx.doi.org/10.1016/j.jnoncrysol.2011.11.010
[3] K. Urban: Phys. Stat. Sol. A 56, 157–168 (1979).
[4] K. Urban: Electron Microsc. 4, 188–195 (1980).
[5] H. Fujita: Hiheikou-zairyo no Riron to Gijyutsu, Seminar text of Japan Institute of Metals, 73–82 (1989). (in Japanese)
[6] W.A. Mckinley, H. Feshbach: Phys Rev., 74, 1759-1763 (1948).
[7] F. Seitz, J.S. Koehler: in: Seitz, F., Tumbull, D. (Eds.) Solid State Physics, Vol. 2, Aca-demic Press, New York, (1956).
[8] J.W. Corbett: in Electron Radiation Damage in Semiconductors and Metals, Academic Press, New York, (1966).
[9] O.S. Oen: Nucl. Instrum. Methods Phys. Res. Sect. B 33, 744–747 (1988).




Review papers for High Voltage Electron Microscopy (Materials Science)

REFERENCES
[1] H. Fujita: J. of Electron Micro. Tech. 3, 243–304 (1986).
[2] H. Fujita.: J. of Electron Micro. Tech. 12, 201–218 (1989).
[3] A. Seeger: J. of Electron Micro. 48, 301-305 (1999).
[4] H. Fujita: Proc. Jpn. Acad., Ser. B, Vol. 81, 141-155 (2005).
[5] H. Mori: J. of Electron Micro., 60, S189-S197 (2001).
[6] H. Yasuda: Kenbikyo, 46, 160-164 (2011). (in Japanese)